Sotaro Nishioka
;
Naoki Sakuyama
;
Ken Obara
;
Akiko T. Saito
;
Mitsuharu Yashima
;
Hidekazu Mukuda
説明:
(abstract)In nuclear magnetic resonance (NMR) spectroscopy, the signal from nuclear spins is extremely weak, and thus, it has been desired to improve the signal-to-noise ratio (SNR). In this study, we designed a compact low-noise probe and peripheral detection system optimized for low-temperature solid-state NMR spectroscopy. We confirmed the operation of a Tx/Rx switch composed of radio frequency microelectromechanical system (RF MEMS) switches and a preamplifier using high electron mobility transistors (HEMTs), which can operate even at 20 K and under a high magnetic field of ∼10 T. In our compactly designed NMR system, we succeeded in sufficiently reducing both thermal noise and external noise. As a result, the 195Pt- and 27Al-NMR measurements for Pt and Al solid powder samples were performed at 12 T and down to 20 K, showing that the SNR was improved by 40%–50% compared with the conventional system. These experiments also provide information on the low temperature properties and operational characteristics of integrated RF electronic devices, such as the RF MEMS switch, the HEMT, and their combined circuit system.
権利情報:
キーワード: NMR, MEMS switch, Cryogenic amplifier, HEMT, Cryogenics
刊行年月日: 2025-10-01
出版者: AIP Publishing
掲載誌:
研究助成金:
原稿種別: 著者最終稿 (Accepted manuscript)
MDR DOI: https://doi.org/10.48505/nims.6126
公開URL: https://doi.org/10.1063/5.0284141
関連資料:
その他の識別子:
連絡先:
更新時刻: 2026-01-15 09:32:30 +0900
MDRでの公開時刻: 2026-01-15 12:22:38 +0900