Yoshikawa, HIdeki
;
Tanuma, Shigeo
;
Isoda, Yukihiro
;
Imai, Motoharu
;
Nishio, Mitsuaki
Description:
(abstract)Electron probe micro analyzer is used for the composition analysis of a wide variety of materials, and its quantification is almost established as a ZAF method. The quantification of MgGe alloy, however, showed an anomalously-high total concentration (ie.120%) of the elements of Mg and Ge. The present analysis of MgGe alloy was not considered to be correct according to the additional fact that the estimating concentration of Mg increase as the increase of accelerating voltage of the electron beam. Because the mass absorption coefficient of Ge for MgKa, is an essential parameter of the ZAF method, we then determined the mass absorption coefficient of Ge for MgKa by using measured X-ray peak intensity ratios of Ge and Mg of MgGe alloy to the standard materials (MgO and Ge). The resulting mass absorption coefficient of Ge for MgKa was 5561±200 cm2/g which were obtained between the electron-beam accelerating voltage of 15kV ~and 30kV. It is smaller than the previous mass absorption coefficient of 7510 cm2/g, which lead the anomalous quantification error. The difference of the mass absorption coefficients must be caused by the difficulty of the measurement of the mass absorption coefficient of Ge for MgKa energy of which is close to the Ge LⅠand LⅡabsorption edges.
Rights:
Creative Commons BY-NC-ND Attribution-NonCommercial-NoDerivs 4.0 International
Keyword: Mg-Ge alloy, electron probe microanalyzer, ZAF, mass absorption coefficient
Date published: 2017-10-05
Publisher: Surface Analysis Society of Japan
Journal:
Funding:
Manuscript type: Not a journal article
MDR DOI:
First published URL: https://doi.org/10.1384/jsa.23.149
Related item:
Other identifier(s):
Contact agent:
Updated at: 2024-06-21 19:04:42 +0900
Published on MDR: 2021-08-14 03:56:51 +0900
| Filename | Size | |||
|---|---|---|---|---|
| Filename |
Vol.23_No.3_03_Nishio.pdf
(Thumbnail)
application/pdf |
Size | 1.04 MB | Detail |