# Fileset

[CardiovascIntervTherapeutics39(2024)428-437.pdf](https://mdr.nims.go.jp/filesets/83677553-4221-434a-88d1-8a612e2b3ef1/download)

## Creator

Sho Torii, [Akiko Yamamoto](https://orcid.org/0000-0002-9182-4886), Ayako Yoshikawa, Linhai Lu, Makoto Sasaki, Shoko Obuchi, Akira Wada, Hideo Tsukamoto, [Gaku Nakazawa](https://orcid.org/0000-0002-6721-1342)

## Rights

This article is licensed under a Creative Commons Attribution
4.0 International License, which permits use, sharing, adaptation,
distribution and reproduction in any medium or format, as long
as you give appropriate credit to the original author(s) and the source,
provide a link to the Creative Commons licence, and indicate if changes
were made. The images or other third party material in this article are
included in the article’s Creative Commons licence, unless indicated
otherwise in a credit line to the material. If material is not included in
the article’s Creative Commons licence and your intended use is not
permitted by statutory regulation or exceeds the permitted use, you will
need to obtain permission directly from the copyright holder. To view a
copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.[Creative Commons BY Attribution 4.0 International](https://creativecommons.org/licenses/by/4.0/)

## Other metadata

[Degradation of a novel magnesium alloy-based bioresorbable coronary scaffold in a swine coronary artery model](https://mdr.nims.go.jp/datasets/6925d88a-77a5-46d6-9e16-6c51639cc83e)

## Fulltext

Degradation of a novel magnesium alloy-based bioresorbable coronary scaffold in a swine coronary artery modelVol:.(1234567890)Cardiovascular Intervention and Therapeutics (2024) 39:428–437https://doi.org/10.1007/s12928-024-01023-3ORIGINAL ARTICLEDegradation of a novel magnesium alloy‑based bioresorbable coronary scaffold in a swine coronary artery modelSho Torii1 · Akiko Yamamoto2 · Ayako Yoshikawa1 · Linhai Lu3 · Makoto Sasaki4 · Shoko Obuchi5 · Akira Wada5 · Hideo Tsukamoto5 · Gaku Nakazawa6 Received: 21 March 2024 / Accepted: 13 June 2024 / Published online: 22 July 2024 © The Author(s) 2024AbstractThe objective of the study is to investigate the safety, feasibility, and degradation profile of a novel Mg alloy-based biore-sorbable coronary scaffold (JFK-PRODUCT BRS) with thin struts (110 μm). Polymer- or Mg alloy-based BRSs have not replaced nondegradable metal stents because of the higher prevalence of scaffold thrombosis and restenosis in clinical practice; these poor clinical outcomes were due to inadequate scaffold designs, including thick struts (more than 150 μm) and their inappropriate degradation processes. Fourteen healthy pigs received 17 JFK-PRODUCT BRSs in the coronary arteries and were sacrificed at 1, 6, 12, 18, and 26 months after implantation. Angiography, optical coherence tomography, microfocus X-ray computed tomography (µCT), scanning electron microscopy with energy-dispersive X-ray spectrometry (SEM–EDX), and histopathological evaluation were performed. The JFK-PRODUCT had a median percent late recoil of 11.28% at 1 month. The µCT observation confirmed that scaffold discontinuity reached 64.8% at 12 months with increased scaffold inner area thereafter, suggesting artery positive remodeling. The inflammation was mild, peaked at 18 months, and decreased thereafter. The SEM–EDX analysis demonstrated gradual degradation of the scaffold with formation of inorganic deposits, presumed to be calcium phosphates. It also revealed the disappearance of calcium phosphates at 26 months, achiev-ing almost complete replacement of the scaffold by biocomponents. The current study demonstrated the safety and feasibility of JFK-PRODUCT with a lower acute recoil rate despite its thin struts. The scaffolds were almost completely disappeared at 26 months after implantation. *  Gaku Nakazawa   gnakazawa@med.kindai.ac.jp1  Department of Cardiology, School of Medicine, Faculty of Medicine, Tokai University, 143 Shimokasuya, Kanagawa 2591193, Japan2  Research Center for Macromolecules and Biomaterials, National Institute for Materials Science, Ibaraki, Japan3  Shanghai Kepan Investment and Management CO., LTD, Shanghai, China4  Faculty of Advanced Science and Technology, Kumamoto University, Kumamoto, Japan5  Japan Medical Device Technology Co., Ltd, Kumamoto, Japan6  Department of Cardiology, Faculty of Medicine, Kindai University, Osaka, Japanhttp://crossmark.crossref.org/dialog/?doi=10.1007/s12928-024-01023-3&domain=pdfhttp://orcid.org/0000-0002-6721-1342429Degradation of a novel magnesium alloy‑based bioresorbable coronary scaffold in a swine coronary…Graphical abstractKeywords  Percutaneous coronary intervention · Bioabsorbable magnesium scaffolds · Preclinical study · Elemental mapping · Controlled degradation profileIntroductionBioresorbable scaffolds (BRSs) have been introduced to counter the well-known limitations of current drug-eluting stents (DESs) in percutaneous coronary artery intervention, which include delayed healing and neoatherosclerosis, lead-ing to restenosis and very late stent thrombosis [1, 2]. As the BRS disappears over time, it clears the above-mentioned issues of the current DESs and provides benefits to leave behind only the native vessel for future treatment strate-gies such as stent/scaffold implantation and coronary artery bypass graft. These expectations are based on several pre-clinical and clinical studies for BRSs which was completely associated with late vessel lumen enlargement and restora-tion of vasomotion [2–4]. Thus, BRSs have the potential to become an important treatment strategy for patients who require coronary revascularization. However, they have not yet replaced DESs because of their various imperfections.The first BRS commercialized was poly(L-lactide)-based Absorb GT-1 (Abbott Vascular, Santa Clara, CA, USA); however, it had a significantly higher incidence of scaffold thrombosis than the Xience DES (Abbott Vascular) in a meta-analysis [5], owing to its thicker strut (157 μm) and extended degradation period (3–5 years) [3, 6]. A Mg alloy-based BRS, Magmaris (DREAMS 2G, Biotronik, Berlin, Germany), was developed and commercialized in Europe [7–9], although this also has a thick strut (150 μm) [7] to maintain a certain level of radial force. The degrada-tion period of Magmaris is shorter than 12 months; however, the late lumen loss was numerically higher than that of the currently available DESs, mainly because of the reduced radial force in the early phase after implantation. In addition, the scaffold leaves amorphous calcium phosphates (substi-tuting Mg alloy backbone) in the artery tissue for more than 2 years [10]. Recently, the DREAMS 3G (Biotronik, Ber-lin, Germany), the third generation of the Magmaris, was developed, which has a thinner strut made of a new Mg–Al alloy [11]. However, a preclinical study demonstrated that the scaffold was fully substituted with calcium phosphate within 12 months, which is similar to that of the earlier gen-eration Magmaris [12].Against this background, a novel BRS (JFK-PRODUCT, Japan Medical Device Technology Co. Ltd., Kumamoto, Japan) was recently developed to overcome the drawbacks of the currently available BRSs. The aim of this study was to evaluate the safety, feasibility, and degradation profile of the JFK-PRODUCT using a porcine coronary artery model for up to 26 months.430  S. Torii et al.MethodsTest deviceThe Mg alloy employed in the JFK-PRODUCT was free of rare earth (RE) elements such as yttrium (Y), dysprosium (Dy), neodium (Nd), and gadolinium (Gd). The alloy surface was passivated by hydrofluoric acid and then coated with three layers: Parylene C, poly (DL-lactide-co-ε-caprolactone), and poly (DL-lactide) [13, 14]. The concentration of Sirolimus was 1.0 µg/mm2. The JFK-PRODUCT consisted of an open cell with a six-crown two-link scaffold with a diameter of 2.5–3 mm and a strut thickness of 110 μm [15]. (Supplemen-tal Figure S1).Experimental modelThe study protocol was reviewed and approved by the Institu-tional Animal Care and Use Committee of AccelLAB (Que-bec, Canada) and Gateway Medical Innovation Center (Shang-hai, China). All the animals were received aspirin (100 mg) and clopidogrel (75 mg), at least 72 h prior to intervention and continuing daily until euthanasia. The medications were crushed and mixed with their food or with an appropriate carrier. A total of 14 swine received JFK-PRODUCT in 17 coronary arteries and were sacrificed at 28 days [1 month (M), n = 3], 180 days (6 M, n = 4), 365 days (12 M, n = 4), 540 days (18 M, n = 4), and 780 days (26 M, n = 2) after scaf-fold implantation (Supplemental Table S1). After pretreatment angiography, the scaffold was deployed to achieve a balloon-to-artery ratio from 1.05 to 1.10:1 based on Quantitative coro-nary angiography (QCA) measurements. Detailed procedures are provided in the supplemental material.Quantitative coronary angiography analysisQCA was performed using the Medis QAngio® XA 7.3 sys-tem (Medias, Leiden, the Netherlands) as previously reported. For angiograms recorded at orthogonal angles in the coronary arteries, the mean values of parameters measured or calculated from the two angles were reported. The detailed method for QCA analysis is described in the supplemental material.Optical coherence tomography analysisOptical coherence tomography imaging was performed using the ILUMIEN or ILUMIEN Optics imaging system (Abbott Vascular, Santa Clara, CA, USA). After scaffold implantation, OCT analysis was performed for all treated vessels. The vessel patency, scaffold and strut adherence to the vessel wall, and constructive change of the scaffold struts were evaluated on the basis of the OCT images of implanted vessel sections. OCT was also performed before euthanasia. The detailed method for OCT analysis is described in the supplemental material.Scaffold harvestThe scaffolds were harvested at designated time points. After nitroglycerin was delivered to the treated arteries, angiog-raphy and OCT were performed. The hearts were harvested and perfused with lactated Ringer’s solution, and fixation perfusion was performed using neutral buffered formalin (NBF). Treated vessels were exposed to NBF for approxi-mately 24 ± 4 h before further processing.Microfocus X‑ray computed tomography (µCT) analysisBefore histological processing, the degradation of the scaf-fold and strut discontinuity were analyzed based on µCT observation (SMX-90CT, Shimadzu, Kyoto, Japan, or XT H225ST, Nikon, Tokyo, Japan) under optimal conditions for each µCT system. The remaining volume of the scaffold was estimated on the basis of histogram data using a peak fitting program [16]. Strut discontinuity was manually ana-lyzed on the 3D images visualized using image analysis soft-ware (VGStudio MAX3.0, Volume Graphics, Heidelberg, Germany). The scaffold inner area was measured on 2D sectional images prepared at approximately 3-mm intervals from the proximal end of the scaffold. A detailed method for µCT analysis is shown in the supplemental material.Histological preparation and assessmentFor light microscopy, the treated vessel segments were dehy-drated in a graded series of ethanol and embedded in Spur’s epoxy resin. After polymerization, the proximal, middle, and distal segments, approximately 2 to 3 mm, were sawed from the devices as previously described [17–19]. Sections from the implants were sliced by a rotary microtome at 4 to 6 µm, mounted, and stained with hematoxylin and eosin (H&E), Elastin Trichrome, and modified Movat Pentachrome stains.Morphometric analysis and histological assessment of scaffolded sections were performed as previously described [17, 20, 21]. The detailed method for morphometric analy-sis is described in the supplemental material. Histological assessments, such as vessel injury and inflammation score, were semi-quantitatively scored for each section. Struts with 431Degradation of a novel magnesium alloy‑based bioresorbable coronary scaffold in a swine coronary…fibrin, > 10 inflammatory cells, giant cells, and calcification were counted and expressed as percentages as previously described [17, 20, 21].Scanning electron microscopy with energy‑dispersive X‑ray spectrometry (SEM–EDX) analysisObservation of the cross-sectional morphologies and ele-ment analysis were performed by field emission-scanning electron microscopy (FE-SEM; JSM-7100F JEOL, Tokyo, Japan) equipped with energy-dispersive X-ray spectroscopy (EDX; JED-2300F, JEOL). The samples were sputter-coated with pure gold (SC-701 AT, Sanyu Electron, Tokyo, Japan) and observed at 15 kV. Additionally, elemental mapping analysis using EDX was performed to characterize the deg-radation layer on the implanted scaffold. EDX mapping of six elements (C, Mg, Ca, P, F, Cl) was performed at an accel-erating voltage of 15 keV with a 120 μm aperture size at 200-fold magnification.Statistical analysisContinuous variables with a normal distribution are expressed as the mean ± SD, while variables with a non-normal distribution are expressed as medians (interquartile range [IQR]). Categorical variables were expressed as counts and percentages, and the chi-square test or Fisher’s exact test was used for comparison. The Kruskal–Wallis test was employed to examine the differences among six independent timepoints due to the non-normal distribution of the data. A p value < 0.05 was considered statistically significant. SAS system Release 9.2 (SAS Institute Inc.) and JMP software (version 13.0, Cary, NC) were used for statistical analyses.ResultsQuantitative coronary angiography and optical coherence tomography analysisAcute recoil was not observed in any of the implanted scaf-folds, and all animals survived the scheduled phases of the study. Angiography demonstrated thrombolysis in myocar-dial infarction (TIMI) 3 flow in all scaffolded vessels, with no luminal thrombus, vessel dissections, or aneurysms. The QCA analysis indicated that the median value of the diam-eter stenosis at 1 M and 3 M was 26.89 [interquartile range (IQR) 23.79 to 29.99] % and 19.42 (IQR: 11.48 to 29.48) %, respectively. The diameter stenosis fell by half at 12 M [8.72 (IQR: 6.87 to 9.53) %], which was followed by lumen enlargement [− 2.94 (IQR: − 5.04 to − 0.83) %] at 26 M fol-low-up (Table 1). A similar tendency was observed in the results of OCT analysis (Supplemental Figure S2). The late recoil at 1 M was 11.28 (IQR: 6.92 to 14.10) %, which then decreased to − 37.55 (IQR: − 46.57 to − 34.59) % at 12 M (Table 2). Serial OCT evaluation of one animal with two scaffolds in the left circumflex and right coronary arteries up to 26 M demonstrated a gradual increase in the lumen area 3 M after implantation, which peaked at 26 M, suggesting positive arterial remodeling (Fig. 1).Micro‑CT analysisThe representative 3D-µCT images of the JFK-PRODUCT demonstrate the progress of scaffold degradation with an increase in the implantation period (Fig. 2). At 1 M, each strut of the scaffold maintained its original width and thick-ness with few discontinuities. White dotty precipitates were observed on the strut surface, indicating less radiolucent substances than the Mg alloy, such as insoluble salts con-taining Ca and P. At 6 M, the width and thickness of the strut Table 1   The results of quantitative coronary angiography (QCA) after the JFK-PRODUCTS implantationFU follow-up1 M (n = 3) 6 M (n = 4) 12 M (n = 4) 18 M (n = 4) 26 M (n = 2)Mean Lumen Diameter Pre (mm) 2.75 ± 0.02 2.55 ± 0.12 2.50 ± 0.03 2.68 ± 0.16 2.72 ± 0.12Balloon to Artery Ratio 1.10 ± 0.01 1.06 ± 0.03 1.09 ± 0.01 1.05 ± 0.06 1.09 ± 0.02Device to Artery Ratio 1.08 ± 0.01 1.06 ± 0.05 1.07 ± 0.03 1.02 ± 0.03 1.07 ± 0.01Minimal Lumen Diameter Post (mm)2.83 ± 0.11 2.60 ± 0.26 2.37 ± 0.12 3.02 ± 0.11 2.80 ± 0.14Minimal Lumen Diameter at FU (mm)1.93 ± 0.11 2.37 ± 0.14 2.37 ± 0.09 3.15 ± 0.36 2.80 ± 0.28Diameter Stenosis at FU (%) 26.89 (23.79–29.99) 14.62 (9.77–18.02) 8.72 (6.87–9.53) 9.06 (−2.16–13.09) −2.94 (−5.04–−0.83)Late Loss at FU (mm) 0.97 (0.94–1.01) 0.25 (0.06–0.42) 0.00 (−0.08–0.08) 0.20 (−0.30–0.29) 0.00 (−0.05–0.05)432  S. Torii et al.appeared to be wider and thicker than those at 1 M, with more uneven deposition of the less radiolucent precipitates on its surface. Some parts of the strut were observed to be darker in color, indicating its substitution with more radio-lucent substances such as MgO, Mg(OH)2, and MgCO3, suggesting local degradation of the struts. At 12 M, some struts completely disappeared, while others were still in the process of degradation. Other remaining parts of the struts were completely substituted by less radiolucent substances. At 18 M, almost all crowns were dismantled and only the less radiolucent substances remained.Figure 3 displays the degradation profiles of the Mg alloy scaffold based on the µCT histogram analysis. As shown in Fig. 3A, we assume constant and isotropic deg-radation of a square-shaped strut. As the total length of the strut is much longer than its thickness and width, the reduction ratio of the strut length by degradation at both ends is negligible compared to that of the strut thickness and width. Therefore, the remaining scaffold volume can be estimated based on the cross-sectional area of the strut. Supposing the initial strut thickness and width as 0.1 mm, we can simulate the remaining volume ratio of the scaffold, as shown in Fig. 3B (details of the calcula-tion is provided in the supplemental material). The dotted line indicates the simulation assuming a 5% decrease in the cross-sectional area of the strut at 1 M, whereas the dashed and dashed-and-dotted lines indicate a 7.5% and 10% decrease at 1 M, respectively. The actual remaining volume ratio based on the µCT images is also plotted in Fig. 3B with the approximated curve indicated as a solid line, which is consistent with the simulation of a 7.5% decrease at 1 M. This agreement indicates that the degra-dation behavior of the JFK-PRODUCT follows the simple hypothesis of constant and isotropic degradation. Based on this simulation, the 50% and complete degradation periods of the JFK-PRODUCT are estimated to be 7 and 23.5 M, respectively.Table 2   Results of optical coherence tomography (OCT) after the JFK-PRODUCTS implantationFU follow-up1 M (n = 3) 6 M (n = 4) 12 M (n = 4) 18 M (n = 2*) 26 M (n = 2)Mean Scaffold Diameter Post (mm)3.06 ± 0.10 2.97 ± 0.18 2.79 ± 0.11 3.02 ± 0.11 2.92 ± 0.14Mean Scaffold Diameter Final (mm)2.90 ± 0.08 2.84 ± 0.06 3.32 ± 0.17 NA NAMean Inner Device Area Post (mm2)7.40 ± 0.46 6.98 ± 0.82 6.12 ± 0.50 7.19 ± 0.53 6.72 ± 0.65Mean Inner Device Area at FU (mm2)6.61 ± 0.37 6.34 ± 0.24 8.76 ± 0.87 7.91 ± 1.74 NALuminal Area at FU (mm2) 4.40 ± 0.88 4.84 ± 0.09 6.55 ± 0.78 7.19 ± 1.41 9.16 ± 0.82Mean Neointimal Area (mm2) 2.21 ± 0.53 1.50 ± 0.18 2.21 ± 0.20 1.87 ± 0.56 NALate Recoil (%) 11.28 (6.92–14.10) 14.43 (1.69–17.03) −37.55 (−46.57–−34.59) −8.47 (−30.40–13.45) NAArea Stenosis (%) 27.82 (27.81–36.60) 23.00 (22.53–24.43) 25.24 (24.22–26.39) 23.57 (22.37–24.77) NAFig. 1   Serial OCT analysis of the JFK-PRODUCT. The data was obtained from one porcine body with two devices implanted into the left circumflex (LCX) and right coronary artery (RCA). The OCT observation was performed at every 2–3 months (M) up to 26 M. The scaffold area was not able to be measured at 24 and 26 M of implan-tation due to degradation of the scaffoldFig. 2   Micro-CT images of the JFK-PRODUCT in the porcine coro-nary arteries. The left five images are representative images for the whole scaffolds at different implantation periods. The right four images are magnified images showing the progress of the strut deg-radation433Degradation of a novel magnesium alloy‑based bioresorbable coronary scaffold in a swine coronary…Fig. 3   Degradation property of the JFK-PRODUCT based on the µCT images. A Schematic explanation of the isotropic degradation of a square-shaped strut, which is employed for the simulation of the degrada-tion profile of a scaffold. B Remaining volume ratio of the JFK-PRODUCT displayed with the simulated profiles suppos-ing a 5%, 7.5%, and 10% loss in the cross-section of the strut at 28 days of implantation. C The scaffold discontinuity ratio plotted against the implantation period. D The scaffold inner area plotted against the scaffold discontinuity ratioFig. 4   a Histopathological observations on the JFK-PRODUCT and vessel seg-ments. Hematoxylin and eosin (H&E) staining images at 1, 3, 6, 12, 18, and 26 months (M) of implantation. B Histologic scores of the JFK-PRODUCT. b A: Inflammation score, B: strut with giant cells, and C: strut with calcification434  S. Torii et al.The scaffold discontinuity ratio is plotted against the implantation time in Fig. 3C. The scaffold discontinu-ity ratio was 1.6% at 1 M, which increased to 46.6% at 6 M, 64.8% at 12 M, and finally reached 97.8% at 18 M. These data indicate a loss of structural integrity at around 6–18 M, ahead of the complete resorption of the Mg alloy backbone at 26 M. Figure 3D shows a plot of the relation-ship between the scaffold inner area and the discontinuity ratio. When the scaffold discontinuity ratio was < 50%, which corresponded to an implantation period of 1–6 M, the scaffold inner area slightly decreased; however, when the scaffold discontinuity ratio was > 50%, which corre-sponded to an implantation period of > 12 M, the scaffold inner area increased even greater than that at the 1 M. This suggests that remodeling of the artery wall accompanies the loss of scaffold integrity.Histologic analysisAll Mg alloy scaffolds were well apposed to the vessel wall, with complete endothelialization at all time points. The scaffolds were almost invisible histologically at 26 M (Fig. 4A). The inflammatory score peaked at 18 M and decreased thereafter at 26 M (Fig. 4B, Table 3). The injury score peaked at 18 M and decreased thereafter with moder-ate medial injury beyond 12 M. At all-time points, fibrin was rarely observed around the struts. The median value of percent struts with giant cells at 6 M was 0.0 (IQR: 0.0 to 6.2) %, which peaked at 18 M [47.1 (IQR: 38.7 to 58.9) %]. The median value of percent struts with calcification at 1 M was 61.1 [IQR: (56.3 to 83.3)] %, which peaked at 6 to 12 M [82.7 (IQR: 73.4 to 95.9) % and 81.2 (IQR: 71.4 to 91.2) %, respectively], and decreased thereafter up to 26 M.SEM–EDX analysisSEM–EDX evaluation of the struts showed that the detected elements varied with the period of implantation (Fig. 5A). At 1 M, only Mg was detected in the strut center, whereas Ca and P were observed in the strut periphery. At 6 M, the struts partially disappeared, and Ca and P were observed in the strut region, indicating the degradation of the Mg alloy backbone. At 18 M, the proportion of carbon (C) in the strut region further increased, whereas those of Ca and P decreased with less detection of Mg. At 26 M, most of the strut region was replaced by C derived from the vascular tissue. These observations indicate a synchro-nous transition of Ca and P in the strut region, suggest-ing the formation of calcium phosphate associated with Mg alloy corrosion and its resorption in the later phase of degradation. Further observation of the strut regions at 26 M revealed a partial loss of Cl, which constitutes Table 3   Histologic Score of the JFK-PRODUCTS1 M (n = 3) 6 M (n = 4) 12 M (n = 4) 18 M (n = 4) 26 M (n = 2) pMorphometric analysisExternal elastic lamina area (mm2)6.9 (6.0–9.5) 8.6 (6.5–13.7) 6.3 (1.8–7.6) 8.3 (5.0–11.6) 7.5 (6.5–8.4) 0.3Scaffold area (mm2) 5.9 (4.7–6.8) 7.3 (4.9–11.5) 5.6 (5.0–6.7) 7.6 (4.1–10.2) 6.1 (5.1–7.1) 0.4Lumen area (mm2) 3.9 (2.8–4.6) 4.3 (2.4–6.9) 3.3 (2.5–4.9) 4.8 (1.7–6.9) 3.1 (2.8–3.4) 0.4Underlying plaque area (mm2)2.0 (1.9–2.2) 2.6 (2.0–5.6) 2.3 (1.8–2.4) 2.8 (2.3–3.3) 3.0 (2.3–3.7) 0.2Media area (mm2) 1.3 (1.1–2.7) 1.7 (1.1–2.2) 0.9 (0.7–0.9) 1.0 (0.6–1.4) 1.4 (1.4–1.4) 0.1Maximum percent stenosis (%)34.6 (32.9–40.9) 51.7 (28.0–54.3) 41.3 (28.6–50.1) 36.8 (32.6–62.0) 48.6 (45.3–51.9) 0.4Mean neointimal thickness (mm)0.3 (0.2–0.4) 0.3 (0.2–0.5) 0.2 (0.2–0.3) 0.3 (0.3–0.3) 0.4 (0.3–0.5) 0.4Pathologic analysis Injury Score 1.6 (1.5–1.8) 1.9 (1.8–2.0) 2.2 (2.0–2.3) 2.3 (1.8–2.5) 2.2 (2.0–2.5) 0.02Strut with Fibrin (%) 0.0 (0.0–4.2) 0.0 (0.0–0.0) 0.0 (0.0–0.0) 4.1 (0.0–13.9) 0.0 (0.0–0.0) 0.4Fibrin score 0.0 (0.0–0.0) 0.0 (0.0–0.0) 0.0 (0.0–0.0) 0.2 (0.0–0.6) 0.0 (0.0–0.0) 1.0Inflammation score 0.0 (0.0–0.0) 0.8 (0.3–1.6) 1.3 (0.8–1.8) 2.7 (2.1–3.0) 1.8 (1.0–2.7) 0.03Strut with inflamma-tory cells (%)0.0 (0.0–0.0) 17.8 (3.7–31.2) 24.6 (10.4–38.2) 66.4 (52.7–88.5) 35.9 (17.8–54.1) 0.04Strut with giant cells (%)0.0 (0.0–0.0) 0.0 (0.0–6.2) 10.2 (4.2–17.6) 47.1 (38.7–58.9) 18.7 (5.6–31.8) 0.03Strut with calcifica-tion (%)61.1 (56.3–83.3) 82.7 (73.4–95.9) 81.2 (71.4–91.2) 70.5 (26.0–81.5) 59.3 (18.5–100.0) 0.6435Degradation of a novel magnesium alloy‑based bioresorbable coronary scaffold in a swine coronary…Parylene C (Fig. 5B), indicating the degradation and dis-appearance of the Parylene C coating in porcine artery tissue. These findings support the complete resorption of the JFK-PRODUCT, including calcium phosphate and the coating layers on its strut surface.DiscussionIn the current study, we performed a comprehensive analy-sis of the JFK-PRODUCT using angiography, OCT, µCT, histopathology, and SEM–EDX analysis in healthy porcine coronary arteries at multiple time points ranging from 1 to 26 M. The detailed characteristics of the JFK-PRODUCT have been reported previously [13, 14]. The JFK-PROD-UCT uses thin struts (110 µm) and a RE-free Mg alloy with fluorine treatment followed by a Parylene C coating, which allows the scaffold to be completely resorbed in ~ 2 years, with no remaining calcium phosphate in the tissue. The prin-cipal findings of this study can be summarized as follows: 1) angiography and OCT analysis confirmed vascular patency without acute recoil throughout the study in all the scaffolds implanted; 2) vessel enlargement was observed at 6 M, and this trend was further augmented beyond 12 M; 3) µCT, SEM–EDX, and histopathological analysis confirmed partial substitution of the Mg alloy strut by insoluble substances containing Ca and P at 6 M as an initial phase of degra-dation, while further degradation with strut discontinuity was observed at 12 M, followed by complete resorption at 26 M; and 4) the inflammatory response increased from 6 M, peaked at 18 M, and decreased thereafter.The complete resorption of the scaffold, including cal-cium phosphate, was confirmed in the current animal study. As shown in the µCT images and SEM–EDX analysis, depo-sition of Ca and P was observed at 1 M, mainly outside the coating layer. Substitution of the Mg alloy backbone by Fig. 5   A Co-registration of the representative images of histol-ogy, OCT, µCT, and SEM–EDX at different implantation periods. B. SEM–EDX images of the strut with distributions of Cl, Mg, and C. The red arrows on the magnified images of the Cl distribution at 26 M indicate the disruption of the Parylene C layer on the strut surface436  S. Torii et al.calcium phosphates partially occurred at 6 M and longer implantation periods, which is a similar process to the cur-rently available Mg alloy BRS (Magmaris) [10]. The cal-cium phosphate deposition of the JFK-PRODUCT peaked at 6–12 M and disappeared at 26 M. Histological evaluation, however, shows calcification in 59.3% of struts at 26 months (Table 3), which may represent a transitional state from cal-cium to carbon. In contrast, calcium phosphate deposition associated with the Magmaris scaffold remained even 2 years after implantation in micro CT analysis [10]. This can be attributed to the difference in Mg alloy backbones, in that the JFK-PRODUCT employs an original RE-free alloy (Mg-1.45%Zn-0.14%Mn) [14], whereas the Magmaris employs an alloy with RE elements such as Y (4 wt%), Dy (0.5 wt%), Nd (2 wt%), and Gd (0.5 wt%) [22]. The DREAMS 3G, the latest version of the Magmaris-BRS, employs an RE-free, Mg–6.25% Al alloy [12], but a recent preclinical study demonstrated its complete substitution by calcium phosphate (containing Al) at 12 M after implantation into porcine coro-nary arteries [12]. The JFK-PRODUCT is a unique Mg alloy BRS that achieves almost complete resorption, leaving no calcium phosphate at 26 M after implantation.In the QCA results of the JFK-PRODUCT, the diameter stenosis was approximately 25% at 1 and 3 M, and it was significantly reduced to approximately one third at the 12 M follow-up. Comparing these results to those of a previous Magmaris preclinical study [23], the findings favor the Mag-maris at the beginning of the implantation period, before reversing after 12 M. This may be attributed to the influ-ence of the polymer coating on the BRSs at the early stages of implantation, whereas at the later stage, the degradation of the polymer itself and other factors, such as moderate medial damage due to moderate inflammation secondary to the degradation process, may contribute to the improve-ment in stenosis and vessel enlargement. This trend was also observed in the recoil rate by OCT analysis. This tendency is consistent with the time course of negative and positive vascular remodeling, which is an inherent vascular behavior in porcine coronary models, as revealed in previous studies [24–26]. Further studies in the clinical settings are needed to evaluate the relationship between vessel enlargement and degradation of the polymer coating.An overview of the degradation process of the JFK-PRODUCT is illustrated in Graphical Abstract, with serial changes observed in the scaffold/lumen area obtained by OCT, the remaining scaffold volume ratio, and the scaffold discontinuity ratio obtained by µCT. The results revealed an initial reduction in the scaffold/luminal area at 3–6 M, followed by a secondary increase at 18 and 26 M, indicat-ing positive remodeling due to vessel uncaging by the loss of scaffold integrity. This enlargement timing was slightly slower than that of the Magmaris (12 M) [23], which may be attributed to the slower degradation of the JFK-PRODUCT than the Magmaris. The slower controlled degradation of the JFK-PRODUCT may be due to the surface treatments with fluorination and Parylene C coating, which improved the corrosion resistance of the Mg alloy scaffold by in vitro testing [14]. Parylene C is a commonly used surface coating material for medical devices such as stents and pacemakers, with no previously reported hazards and toxicity.Study limitationsThe current study has several limitations that warrant discus-sion. First, the number of scaffolds evaluated was relatively small. Second, this study was conducted using a healthy porcine coronary artery model; therefore, the results cannot consider the effect of atherosclerosis on the degradation of and tissue response to the JFK-PRODUCT. Nevertheless, serial follow-up with detailed evaluation using several imag-ing modalities provides mechanistic and biological insight into the degradation process of the JFK-PRODUCT.ConclusionsIn the current preclinical study using a healthy porcine coronary artery model, a novel Mg alloy-based BRS, JFK-PRODUCT, demonstrated its safety with complete resorp-tion of the Mg alloy backbone, leaving no calcium phos-phates in the tissue at 26 M. Further clinical studies are needed to confirm the findings of this preclinical study.Supplementary Information  The online version contains supplemen-tary material available at https://​doi.​org/​10.​1007/​s12928-​024-​01023-3.Acknowledgements  We wish to express our gratitude to Takafumi Ueno MD, PhD (former professor at Kurume University) and Akira Mochizuki PhD (former professor at Tokai University) for their advice on this research and development. We also thank the Technical Service Coordination Office, Tokai University, for their technical support with the SEM-EDX measurements.Funding  This research was supported by AMED under Grant Number 18pc0101027h0001.Data availability  The data that support the findings of this study are available from the corresponding author upon reasonable request. Due to the sensitive nature of the clinical data, it has not been made publicly available. Researchers who wish to access the data should contact Gaku Nakazawa at [gnakazawa@med.kindai.ac.jp] to discuss the conditions under which the data can be accessed.Declarations Conflict of interest  This study was sponsored by Japan Medical De-vice Technology Co., Ltd. (Kumamoto, Japan) with support by AMED under Grant Number JP18pc0101027. The magnesium alloy used was provided by Fuji Light Metal Co., Ltd. (Kumamoto, Japan). Sho Torii received research grants from Abbot Vascular Japan, Boston Scien-https://doi.org/10.1007/s12928-024-01023-3437Degradation of a novel magnesium alloy‑based bioresorbable coronary scaffold in a swine coronary…tific Japan, Medtronic, and honoraria from Boston Scientific Japan. Gaku Nakazawa is a consultant for Boston Scientific, Abbott Vascular, Terumo Corp., and Japan Medical Device Technology Co., Ltd.; He received research grants from Boston Scientific, Abbott Vascular, and Terumo. Makoto Sasaki, Shoko Obuchi, Akira Wada, and Hideo Tsu-kamoto are employees of Japan Medical Device Technology Co., Ltd.Generative AI and AI‑assisted technologies in the writing process  Gen-erative AI and AI-assisted technologies were not used in the current manuscript.Open Access  This article is licensed under a Creative Commons Attri-bution 4.0 International License, which permits use, sharing, adapta-tion, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.References  1.  Torii S, Jinnouchi H, Sakamoto A, et al. Drug-eluting coronary stents: insights from preclinical and pathology studies. Nat Rev Cardiol. 2020;17:37–51.  2.  Jinnouchi H, Torii S, Sakamoto A, Kolodgie FD, Virmani R, Finn AV. Fully bioresorbable vascular scaffolds: lessons learned and future directions. Nat Rev Cardiol. 2019;16:286–304.  3.  Otsuka F, Pacheco E, Perkins LE, et al. Long-term safety of an everolimus-eluting bioresorbable vascular scaffold and the cobalt-chromium XIENCE V stent in a porcine coronary artery model. Circ Cardiovasc Interv. 2014;7:330–42.  4.  Yahagi K, Yang Y, Torii S, et al. Comparison of a Drug-Free Early Programmed Dismantling PDLLA Bioresorbable Scaffold and a Metallic Stent in a Porcine Coronary Artery Model at 3-Year Follow-Up. J Am Heart Assoc. 2017;2017:6.  5.  Ali ZA, Serruys PW, Kimura T, et al. 2-year outcomes with the Absorb bioresorbable scaffold for treatment of coronary artery disease: a systematic review and meta-analysis of seven ran-domised trials with an individual patient data substudy. Lancet. 2017;390:760–72.  6.  Sotomi Y, Onuma Y, Collet C, et  al. Bioresorbable Scaf-fold: The Emerging Reality and Future Directions. Circ Res. 2017;120:1341–52.  7.  Haude M, Ince H, Abizaid A, et al. Sustained safety and perfor-mance of the second-generation drug-eluting absorbable metal scaffold in patients with de novo coronary lesions: 12-month clinical results and angiographic findings of the BIOSOLVE-II first-in-man trial. Eur Heart J. 2016;37:2701–9.  8.  Lipinski MJ, Acampado E, Cheng Q, et al. Comparison of acute thrombogenicity for magnesium versus stainless steel stents in a porcine arteriovenous shunt model. EuroIntervention. 2019;14:1420–7.  9.  Waksman R, Lipinski MJ, Acampado E, et al. Comparison of Acute Thrombogenicity for Metallic and Polymeric Bioabsorbable Scaffolds: Magmaris Versus Absorb in a Porcine Arteriovenous Shunt Model. Circ Cardiovasc Interv. 2017;2017:10. 10.  Joner M, Ruppelt P, Zumstein P, et al. Preclinical evaluation of degradation kinetics and elemental mapping of first- and second-generation bioresorbable magnesium scaffolds. EuroIn-tervention. 2018;14:e1040–8. 11.  Haude M, Wlodarczak A, van der Schaaf RJ, et al. A new resorb-able magnesium scaffold for de novo coronary lesions (DREAMS 3): one-year results of the BIOMAG-I first-in-human study. Euro-Intervention. 2023;19:e414–22. 12.  Seguchi M, Baumann-Zumstein P, Fubel A, et al. Preclinical eval-uation of the degradation kinetics of third-generation resorbable magnesium scaffolds. EuroIntervention. 2023;19:e167–75. 13.  Xu W, Yagoshi K, Koga Y, Sasaki M, Niidome T. Optimized poly-mer coating for magnesium alloy-based bioresorbable scaffolds for long-lasting drug release and corrosion resistance. Colloids Surf B Biointerfaces. 2018;163:100–6. 14.  Sasaki M, Xu W, Koga Y, et al. Effect of Parylene C on the Corrosion Resistance of Bioresorbable Cardiovascular Stents Made of Magnesium Alloy “Original ZM10.” Materials (Basel). 2022;2022:15. 15.  Shimizu I, Wada A, Sasaki M. A Study on Designing Balloon Expandable Magnesium Alloy Stent for Optimization of Mechani-cal Characteristics. Proceedings. 2018;2:523. 16.  Wojdyr M. Fityk : a general-purpose peak fitting program. J Appl Crystallogr. 2010;43:1126–8. 17.  Nakazawa G, Shinke T, Ijichi T, et al. Comparison of vascular response between durable and biodegradable polymer-based drug-eluting stents in a porcine coronary artery model. EuroInterven-tion. 2014;10:717–23. 18.  Nakamura N, Torii S, Aihara K, et al. Poor Below Knee Run-off Impacts Femoropopliteal Stent Failure and Fluoropolymer Antithrombotic Effect in Healthy Swine Model. Eur J Vasc Endo-vasc Surg. 2023;66:722–9. 19.  Nagamatsu H, Torii S, Aihara K, et al. Histological evaluation of vascular changes after excimer laser angioplasty for neointimal formation after bare-metal stent implantation in rabbit iliac arter-ies. Cardiovasc Interv Ther. 2023;38:223–30. 20.  Torii S, Yahagi K, Mori H, et al. Biologic Drug Effect and Par-ticulate Embolization of Drug-Eluting Stents versus Drug-Coated Balloons in Healthy Swine Femoropopliteal Arteries. J Vasc Interv Radiol. 2018;29:1041-1049.e3. 21.  Torii S, Yahagi K, Mori H, et al. Safety of Zilver PTX Drug-Elut-ing Stent Implantation Following Drug-Coated Balloon Dilation in a Healthy Swine Model. J Endovasc Ther. 2018;25:118–26. 22.  Menze R, Wittchow E. In vitro and in vivo evaluation of a novel bioresorbable magnesium scaffold with different surface modifica-tions. J Biomed Mater Res B Appl Biomater. 2021;109:1292–302. 23.  Waksman R, Zumstein P, Pritsch M, et al. Second-generation magnesium scaffold Magmaris: device design and preclinical evaluation in a porcine coronary artery model. EuroIntervention. 2017;13:440–9. 24.  Lipinski MJ, Escarcega RO, Lhermusier T, Waksman R. The effects of novel, bioresorbable scaffolds on coronary vascular pathophysiology. J Cardiovasc Transl Res. 2014;7:413–25. 25.  Maeng M, Jensen LO, Falk E, Andersen HR, Thuesen L. Negative vascular remodelling after implantation of bioabsorbable mag-nesium alloy stents in porcine coronary arteries: a randomised comparison with bare-metal and sirolimus-eluting stents. Heart. 2009;95:241–6. 26.  Strandberg E, Zeltinger J, Schulz DG, Kaluza GL. Late positive remodeling and late lumen gain contribute to vascular restoration by a non-drug eluting bioresorbable scaffold: a four-year intravas-cular ultrasound study in normal porcine coronary arteries. Circ Cardiovasc Interv. 2012;5:39–46.Publisher's Note  Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.http://creativecommons.org/licenses/by/4.0/ Degradation of a novel magnesium alloy-based bioresorbable coronary scaffold in a swine coronary artery model Abstract Graphical abstract Introduction Methods Test device Experimental model Quantitative coronary angiography analysis Optical coherence tomography analysis Scaffold harvest Microfocus X-ray computed tomography (µCT) analysis Histological preparation and assessment Scanning electron microscopy with energy-dispersive X-ray spectrometry (SEM–EDX) analysis Statistical analysis Results Quantitative coronary angiography and optical coherence tomography analysis Micro-CT analysis Histologic analysis SEM–EDX analysis Discussion Study limitations Conclusions Acknowledgements  References